SEVOFLURANE TANK DATEX OHMEDA SEVOTEC 5

SEVOFLURANE TANK DATEX OHMEDA SEVOTEC 5

Fr892 (CHF)

Beschreibung

SPECIFICATION :
CALIBRATION
The vaporizers are calibrated at 21°C using an oxygen carrier gas at a flow rate of 5 liters / minute and are temperature, flow and pressure compensated within the specified operating range.
GAS FLOW RESISTANCE WITH VAPORIZER SWITCHED OFF
Resistance Vaporizer Setting Flow Temperature Temperature
10 -15 cm wg ON + 0% 5 liters / minute, oxygen 21 ± 2 ° C
LIQUID CAPACITY
The amount of liquid required to fully charge a vaporizer with dry wicks = 300 milliliters (nominal).
The amount retained by the wick system = 75 millilitres (nominal).
5% vaporizer
The amount of liquid needed to fill the vaporizer from minimum to maximum = 170 milliliters (nominal).
8% vaporizer (Sevotec)
The quantity of liquid needed to fill the vaporizer from minimum to maximum = 137 milliliters (nominal).
WEIGHT AND DIMENSIONS
ENVIRONMENTAL CONDITIONS
Operating temperature range = 18°C to 35°C
Storage temperature range = -40°C to +65°C
PERFORMANCE
Accuracy at 5 liters / minute O2 at 21°C ± 2°C:
Vaporizer 5% ± 0.25% of delivered agent or
± 15% of dial setting (whichever is greater).
Spray 8% ± 0.4% of delivered agent or
± 15% of dial setting (whichever is greater).

Spezifikationen

ZustandGebraucht
Lagernummer6741
Weight7.0 kg
Height250 mm
Depth164 mm
Width110 mm
NoteSpray packaging must be protected from condensation.